LEAVING YOUR PRACTICE FOR A LITTLE WHILE, MAKE SURE FOLLOW THESE SHUTDOWN PROCEDURES

LEAVING YOUR PRACTICE FOR A LITTLE WHILE, MAKE SURE FOLLOW THESE SHUTDOWN PROCEDURES

Chair Shutdown procedure for Stern Weber and Castellini

  • Run the unit through a cleaning cycle Autosteril / Bioster or 2 min Flush cycle depending on the version you operate on your unit.
  • Remove Water bottle and Drain lines of all water by activating dynamic instrumentation
  • Drain Air condensate located either at the base of the Chair or under the spittoon
  • Run Suction Cleaning Procedure
  • Clean all filters and gently apply O ring grease or Vaseline to all rubber seals
  • DO NOT LEAVE ANY SUCTION, SPITTOON FILTERS, SUCTION ENDS, IN ANY CLEANING SOLUTION, OF ANY KIND, OTHER THAN WATER
  • Drain your spittoon valve by depressing the Drain activation switch inside the spittoon
  • Switch Off the unit on the Chair and on the wall

 

Procedure for Suction Motor shut down Durr and Cattani

  • Run your cleaning product through the Suction lines
  • Run 2 x volume of Clean Water afterwards
  • Run the Suction motor for 10 minutes with only air running through the suction tubes
  • Switch of the Suction Motor and remove the amalgam pot and clean and drain
  • Clean suction filter
  • Refit amalgam pot and filter and gently apply O ring grease or Vaseline to all rubber seals
  • Switch Off the Suction unit

 

Compressor Shutdown procedure

  • Turn Compressor Off
  • Drain the Tank of all Air with Tank Drain Valve
  • Drain the Dryer Column where applicable